Country of Origin | |
Name of Manufacturer | |
Brand Name | |
Particulars | Requirements |
Purpose | RNA extraction suitable for manual extraction of RNA from sputum, oropharyngeal swab, nasopharyngeal swab body fluids and serum/Plasma samples. |
Description | Spin column based
Sample input: Up to 400 microliter Elution volume: More than 30 microliter with RNA ready for Real Time PCR Should not require heating step during extraction Extraction steps should not take more than thirty minutes. The number of collection tubes should be provided in sufficient number
|
Other requirements | The manufacturer should be certified by WHO or USFDA or EUCE or should have USFDA/CDC emergency use authorisation or WHO Emergency Use Listing |
- RNA Extraction Kit Automated
S.N. | Purchaser’s Specifications |
B | RNA Extraction Kit Automated Compatible with (Nanjing Zhongkebio Medical Technology co Ltd., Brand name ZhongkeBio Type model ZK-96) |
Manufacturer | |
Brand | |
Type / Model | |
Country of Origin | |
1 | Description of Function |
1.1 | Extraction of Nucleic acid (RNA) with all accessories and consumables necessary for asked number of tests |